LUNG-PROTECTIVE VENTILATION — adherence ============================================================================ Idea 17. Deterministic: computes predicted body weight from height and sex, derives the tidal volume target, and flags drift above it.
THE TRAP THIS EXISTS TO CATCH ----------------------------- Tidal volume is set per kilogram of PREDICTED body weight, computed from height, not from what the patient weighs. Using actual weight systematically over-ventilates short patients and obese patients — the lung does not grow with body mass.
A 152 cm woman has a predicted weight of 45.5 kg and a 6 mL/kg target of 273 mL. If she weighs 85 kg and someone sets by actual weight, she receives 510 mL — nearly double, delivered every breath, for days.
No arithmetic here is difficult. It is simply not done at 3 AM on twelve patients, which is the entire case for computing it continuously.
Sources: ARDS Network. NEJM 2000;342:1301-1308. ARDSNet PBW equations, standardised: CHEST 2015.
meta
version0.1.0
ownerUNASSIGNED — set to the responsible clinician
last clinical reviewnot set
verified bynot set
predicted body weight
male
base kg50.0
per cm above0.91
reference height cm152.4
female
base kg45.5
per cm above0.91
reference height cm152.4
note
PBW = base + 0.91 x (height_cm - 152.4). Below the reference height the
formula returns a smaller weight, which is correct — but a very short
patient produces a low PBW, so the computed target must be sanity-checked
rather than applied blindly.
minimum plausible kg30.0
targets
tidal volume
target ml per kg6.0
acceptable max ml per kg8.0
critical ml per kg10.0
minimum ml per kg4.0
note
6 mL/kg is the ARDSNet target, reduced stepwise to as low as 4 if needed
to hold plateau pressure at or below 30. Above 8 is drift; above 10 is
the tidal volume the trial was designed to eliminate.
plateau pressure
target max cmh2o30.0
critical cmh2o35.0
noteMeasured on an inspiratory hold. Not the same as peak pressure.
driving pressure
target max cmh2o15.0
critical cmh2o18.0
formulaplateau pressure - PEEP
note
Independently associated with mortality. Not part of the original
ARDSNet protocol; included because it is computed from values already
recorded and is rarely tracked.
peep
minimum cmh2o5.0
noteSet from the ARDSNet PEEP/FiO2 table, which is unit-specific.
checks
- idvt_above_target
nameTidal volume above target
severity warn ml per kg8.0
severity critical ml per kg10.0
actionReduce tidal volume toward 6 mL/kg predicted body weight
- idvt_below_minimum
nameTidal volume below 4 mL/kg
action
Verify this is deliberate. Below 4 mL/kg is used to hold plateau
pressure but risks hypercapnia and derecruitment.
- idplateau_high
namePlateau pressure above 30
actionReduce tidal volume stepwise by 1 mL/kg, to a minimum of 4
- iddriving_pressure_high
nameDriving pressure above 15
actionReview PEEP and tidal volume together
- idpeep_below_minimum
namePEEP below 5
actionReview against the unit's PEEP/FiO2 table
- idset_by_actual_weight
nameTidal volume appears set by actual weight
action
Recalculate against predicted body weight. Flagged when the delivered
volume is close to 6-8 mL/kg of ACTUAL weight while exceeding target for
predicted — the specific error this engine exists to catch.
trigger tolerance ml per kg1.5
- idno_height_recorded
nameHeight not recorded
action
Predicted body weight cannot be computed, so no target exists. Height is
a one-time measurement and is the single input this entire protocol
depends on.
applies to
invasively ventilatedtrue
nivfalse
spontaneous modestrue
note
Applies to any invasively ventilated patient, not only those with ARDS —
the compliance gap is largest in patients without an ARDS label, because
nobody computed a target for them.