ventilators.yaml

Configuration · ventilators.yaml · 181 lines · editable in place

VENTILATOR ADAPTERS ============================================================================ The detection engine reads flow and pressure arrays and does not care which machine produced them. Adapters translate one manufacturer's protocol into that common shape — nothing more.

WHAT AN ADAPTER COSTS --------------------- Roughly a week each: read the protocol document, implement the framing, map the channel identifiers, verify against a real machine. That is the whole friction, and it buys continuous inspection of every breath — about 20,000 a day per ventilated patient, none of which anyone watches.

WHAT MOST HOSPITALS DISCOVER ---------------------------- The port already exists and was never connected. The cable is a null-modem DB9 and the data has been discarded since the machine was installed. Access is often a matter of asking rather than buying.

STATUS ------ implemented parses real device output stub framing declared, not yet written planned documented, no work started

Nothing here is implemented against a physical ventilator. The synthetic adapter exists so the pipeline runs end to end and the detector can be calibrated; it is not a device.

meta
version0.1.0
ownerUNASSIGNED — set to the responsible biomedical engineer
note
Serial parameters below are from manufacturer documentation and
open-source capture projects. Confirm against the specific model before
connecting anything.
common channels
- idflow
Every adapter maps into these. An adapter that cannot supply a channel reports it as absent rather than substituting a value.
unitL/min
hz50
requiredtrue
noteInspiration positive, expiration negative. Asynchrony lives here.
- idpressure
unitcmH2O
hz50
requiredtrue
- idvolume
unitmL
hz50
requiredfalse
- idco2
unitmmHg
hz50
requiredfalse
common settings
tidal_volume_ml
respiratory_rate_set
respiratory_rate_total
peep
peak_pressure
plateau_pressure
fio2
mode
adapters
- idsynthetic
makerNone — generated
models[]
statusimplemented
transportnone
note
A shape generator for calibrating the detector. Produces flow with a
known answer, so a false positive rate can be measured. Not a device.
- iddraeger_medibus
makerDräger
models[Evita, Evita 2 dura, Evita 4, Evita XL, Evita V300,
Infinity V500, Savina, Savina 300]
statusstub
transportserial
serial
baud19200
databits8
parityeven
stopbits1
flow controlnone
cableDB9 null-modem, COM1 on the rear of the machine
protocolMEDIBUS
protocol note
Request-response with a polled command set. Waveform streaming is a
separate real-time channel from the numeric data. Well documented and
the most widely implemented of the six.
references
"Dräger MEDIBUS protocol documentation"
"VitalSignsCapture (open source) implements MedibusX"
- idhamilton
makerHamilton Medical
models[G5, S1, C6, C3, T1]
statusstub
transportserial
serial
baud9600
databits8
paritynone
stopbits1
protocolHamilton proprietary
protocol note
Newer models also expose an HL7 or network interface, which is
preferable where available — a network feed avoids a cable per bed.
- idgetinge_servo
makerGetinge / Maquet
models[Servo-i, Servo-u, Servo-n, Flow-i]
statusstub
transportserial
serial
baud9600
databits8
paritynone
stopbits1
protocolServo Computer Interface
references
"VitalSignsCapture implements Servo-i, Servo-U, Flow-i"
- idpuritan_bennett
makerMedtronic / Puritan Bennett
models[PB840, PB980]
statusstub
transportserial
serial
baud38400
databits8
paritynone
stopbits1
protocolPB serial
- idmindray
makerMindray
models[SV300, SV600, SV800]
statusstub
transportnetwork
protocolBeneLink / HL7
protocol note
Common in Indian units and priced accordingly. Network rather than
serial on the newer models, which is easier to deploy across a ward.
- idphilips
makerPhilips
models[V60, V680, Trilogy]
statusplanned
transportnetwork
protocolIntelliVue / HL7
protocol note
Mostly non-invasive ventilation, so lower priority for asynchrony work.
deployment
preconditions
What has to be true for an adapter to be useful, stated so it is checked before the week of work rather than after.
"The serial or network port is physically accessible"
"Waveform streaming is enabled in the machine's service settings"
"One cable or network drop per bed, or a serial concentrator"
"Biomedical engineering approval — this is a connection to a life-support device"
read onlytrue
read only note
Adapters read. Nothing writes to a ventilator, and no adapter in this
system will ever accept a command. That is a hard boundary, not a
current limitation.